Development of Heartbeat Anomaly and Hypoxemia Monitoring System based on MySQL and Telegram Mayda Waruni Kasrani, Aswadul Fitri Saiful Rahman, Khairunnisa Nurhandayani, Gusti Nala Fajar
Universitas Balikpapan
Abstract
Everyone needs a good health condition for making productivity. Health conditions can be shown by the performance of the organs in the body such as the heart. The heart that can not work properly has a chance of developing cardiovascular disease. Cardiovascular disease is one of the diseases with the highest death rate in Indonesia, so it has become a big issue in the medical community. A monitoring device for detecting heart rate anomalies and symptoms of hypoxemia based on MySQL and Telegram is built into this research to determine the state of the heart on mobile. This device consists of MAX30100 for reading pulse and Oxygen rate in the blood, and NodeMCU ESP 8266 for processing data from MAX30100 and sending it to the local web server and Telegram. This device was compared with an oximeter made by Mixio for testing the overall system. The reading from the research device and an oximeter by Mixio has a little difference of around 0.85% for BPM error and 0.89% for SpO2. Meanwhile, the data transferred from NodeMCU ESP 8266 to Telegram has a 97% success rate from 100 tests.
Keywords: NodeMCU ESP8266, Telegram, Local Web Server, MAX30100, Database